Our proprietary image reconstruction algorithms are at the heart of the QTI Breast Acoustic CT™ platform, transforming raw ultrasound transmission and reflection data into high-resolution, volumetric images.

Unlike traditional ultrasound systems that rely on 2D image slices and manual probe manipulation, the QT Scanner 2000 captures fully tomographic datasets using a standardized, automated process. During each scan, sound waves are transmitted and received across 180 rotational angles and multiple vertical positions. This produces thousands of projection measurements from both transmission-mode and reflection-mode imaging. What makes QTI unique is how these datasets are processed. Using nonlinear inverse scattering methods, the system reconstructs true 3D volumetric maps of speed of sound and attenuation — quantitative imaging biomarkers that correlate with tissue composition, structure, and pathology.

At the same time, reflection-mode data is spatially compounded across multiple focal depths to dramatically reduce speckle and improve resolution. Importantly, these reflection-mode images are corrected using the transmission-mode speed of sound maps to account for tissue-specific refraction effects, ensuring anatomical fidelity.

The result is a set of co-registered, 3D volumes that currently reflect two imaging modalities in one:

Frontal view of a QT Breast Scan using speed of sound.

Speed of Sound

Providing a quantitative biomarker for tissue and pathology characterization.

Frontal view of a QT Scan using reflection. A section is cut out and zoomed in on with arrows pointing out anatomical structures.

Reflection

Revealing anatomical structures in detailed high resolution.

Composite QT Scan image, color coded to show skin, ducts, glands, connective tissue, and fat.

Tissue Classification Map

Images are combined to generate a respective tissue-color-coded classified image.

Speed of Sound of Breast Tissue Types
(Mean ± SD)

The graph shows the ranges of speed, in meters per second, that different parts of the breast are captured at. Skin ranges between 1529 to 1537 meters per second. Cooper Ligaments 1422 to 1496 meters per second. Ductal Tissue 1560 to 1612 meters per second. Glandular Tissue 1517 to 1567 meters per second. Fat 1418 to 1436 meters per second.
  • Klock JC, Iuanow E, Malik B, Obuchowski NA, Wiskin J, Lenox M. Anatomy-Correlated Breast Imaging and Visual Grading Analysis Using Quantitative Transmission Ultrasound™.
  • Int J Biomed Imaging. 2016;2016:7570406. doi: 10.1155/2016/7570406. Epub 2016 Sep 26. PMID: 27752261; PMCID: PMC5056280.

Volumetric Imaging Pipeline

Our reconstruction pipeline is what makes QTI Breast Acoustic CT™ technology different from conventional ultrasound. It is more aligned with the principles of inherently volumetric imaging modalities, delivering consistent, data-rich, and diagnostic-quality breast imaging without radiation, compression, or contrast agents.