QT Imaging has developed an automated, software-controlled, fully 3D tomographic transmission and reflection imaging system using low-frequency sound waves.
QTI’s Breast Acoustic CT™ scanner is FDA-cleared as a supplemental screening device for breast imaging, including follow-up for inconclusive or suspicious mammogram findings. It may be particularly valuable for women with dense breast tissue where conventional breast imaging modalities may be less effective.

The QTI Breast Acoustic CT™ scanner performs a standardized scan of a whole breast — without using compression, ionizing radiation, or contrast agent injection — and generates true 3D images with remarkable accuracy and precision. The proprietary, AI-powered analysis software uses these volumetric images for breast assessment, including quantifying breast fibroglandular volume (FGV) and total breast volume (TBV) and measuring breast density (or fibroglandular ratio, FGR).
QTI’s Breast Acoustic CT™ scanner has a scan head that contains five ultrasound transducer arrays: a transmission-mode transmitter, a transmission-mode receiver, and a set of three reflection-mode transducers.
Superior Patient Experience
A QTscan is performed, one breast at a time, with the patient laying in a prone position on the examination table with the breast suspended in a warm water bath inside the scan tank. Images are automatically acquired on the pendant breast positioned with the nipple as a point of reference. The transducer arrays rotate about a vertical axis to fully circle the breast in the coronal plane. The arrays are then translated 2 mm vertically, repeating the scanning process until the entire breast is scanned.

Scan Modes
In transmission-mode imaging, the system detects the sound pulse energy that is transmitted through the breast and received on the opposite side. The transmitter emits a plane wave that goes around and through the breast, which is received by a 2048-element (8 x 256) lead magnesium niobate-lead titanate (PMN-PT) receiver array with custom data acquisition electronics. Multiple acquisitions at frequencies ranging from 300 kHz to 2.0 MHz are acquired for 180 angles as the scan head rotates around the breast. The acquired projection data is then reconstructed using nonlinear inverse scattering in 3D. This reconstruction produces a quantitative, high-resolution volume map of both the speed of sound and the attenuation.

The reflection-mode imaging uses three reflection transducers with different focal lengths to extend the overall depth of focus to cover the imaging volume. Each of the three transducers are alternately fired between transmission measurements in a B-mode acquisition. The resulting images are corrected for refraction using the speed of sound map computed in the transmission-mode phase, and then spatially compounded together. This spatial compounding dramatically reduces speckle while maintaining the high-resolution, resulting in images that can be read similar to traditional B-mode images.
The end result of each QTscan is automatically co-registered, 3D volume images of speed of sound, attenuation, and reflection. These image volumes can be viewed in all orthogonal views concurrently (coronal, axial, and sagittal) using the QTviewer software.

Speed of Sound

Reflection

Attenuation
Image Characterization
QTI’s proprietary AI-powered analysis software uses the co-registered, high-resolution, volumetric data to systematically characterize and isolate different tissues types (glands, ducts, fat, skin, or connective tissue) and masses (cysts or masses), while providing instructive anatomical references.